Electromagnetic Compatibility (Emc); General Emc Information; Electromagnetic Emission; Electromagnetic Immunity - Invacare Medley Ergo Soft Tilt User Manual

Accessory
Table of Contents

Advertisement

10 Electromagnetic compatibility (EMC)

10.1 General EMC information

Medical Electrical Equipment needs to be installed and used according to the EMC information in this manual.
This product has been tested and found to comply with EMC limits specified by IEC/EN 60601-1-2 for Class B equipment.
Portable and mobile RF communications equipment can affect the operation of this product.
Other devices may experience interference from even the low levels of electromagnetic emissions permitted by the above standard. To
determine if the emission from this product is causing the interference, run and stop running this product. If the interference with the
other device operation stops, then this product is causing the interference. In such rare cases, interference may be reduced or corrected
by the following:
Reposition, relocate, or increase the separation between the devices.

10.2 Electromagnetic emission

Guidance and manufacturer ́ s declaration
This product is intended for use in the electromagnetic environment specified below. The customer or the user of this product should
assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group I
CISPR 114
RF emissions
Class B
CISPR 114
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations /
flicker emissions IEC
Complies
61000-3-3

10.3 Electromagnetic Immunity

Guidance and manufacturer ́ s declaration
This product is intended for use in the electromagnetic environment specified below. The customer or the user of this product should
assure that it is used in such an environment.
Immunity test
Test / Compliance level
Electrostatic
± 8kV contact
discharge (ESD)
± 2 kV, ± 4 kV, ± 8 kV, 15
IEC 61000-4-2
kV air
± 2 kV for power supply
lines; 100 kHz repetition
Electrostatic
frequency
transient / burst
± 1 kV for input / output
IEC 61000-4-4
lines; 100 kHz repetition
frequency
Surge
± 1 kV line to line
IEC 61000-4-5
± 2 kV line to earth
IVCMAMED221119
Electromagnetic environment – guidance
This product uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
This product is suitable for use in all establishments including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic
material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Electromagnetic compatibility (EMC)
23

Advertisement

Table of Contents
loading

This manual is also suitable for:

Soft tilt

Table of Contents