How To Program The Individual Automated Programs - Invacare Medley Ergo Soft Tilt User Manual

Accessory
Table of Contents

Advertisement

Invacare® Soft Tilt
5.2
How to program the individual
automated programs
P Set time / Start programming mode button
Q Set angle button
R Adjust button Plus (increase values for time and angle)
S Adjust button Minus (decrease values for time and angle))
T Position indicators (LEDs)
Right (top) = right wing or position
Flat (middle) = both wings in horizontal position
Left (bottom) = Left wing or position
When programing the Soft Tilt, first the time period for which the wings are in a elevated or flat position is set, second the angles at which to
elevate and third the starting side of the wings (right, left or flat).This is done for either Program 1 or Program 2.
Step
Function
1
Press and
hold for3
sec.
2
Press
or
3
Press
4
Press
5
Press
6
Press
14
Display
LED
SEL
None
P1
P1
P2
P2
010
P1 or P2
&
Right
001
P1 or P2
&
020
Right
010
P1 or P2
&
Flat
001
P1 or P2
&
020
Flat
Note
Start the programming mode
Choose the program to be set
The time value appears in the display
Adjust the time between 000 and 180 minutes (0, 3,
10-180 in +/- 10 min. intervals)
The time value appears in the display
Adjust the time between 000 and 180 minutes (0, 3,
10-180 in +/- 10 min. intervals)
IVCMAMED221119

Advertisement

Table of Contents
loading

This manual is also suitable for:

Soft tilt

Table of Contents