Emc Information; Electromagnetic Compatibility (Emc) Information - Invacare CS3 User Manual

Cs series medical beds
Hide thumbs Also See for CS3:
Table of Contents

Advertisement

3 EMC INFORMATION

3 EMC Information
3.1

Electromagnetic Compatibility (EMC) Information

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect medical electrical equipment.
WARNINGS
The use of accessories, tranducers, and cables other than those specified by the manufacturer, may result in increased
emissions or decreased immunity of this bed.
This bed should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary,
this bed should be observed to verify normal operation in the configuration in which it will be used.
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The CS Bed is intended for use in the electromagnetic environment specified below. The customer or the user of the CS Bed should assure that it is used
in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Invacare ®CS Series Beds
Compliance
Group 1
The CS Bed uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic
equipment.
Class B
The CS Bed is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Class A
Complies
16
Electromagnetic Environment - Guidance
Part No 1153410

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cs5Cs9Cs7Cs9fx600

Table of Contents