The hand control will not work when a program is
active.
Status indicators
Active Program Display
AUTO
AU
Porgram 1
P1
Program 2
P2
Timer
P1/XX and
P2/XX

4.7.1 Auto program sequence

Starting from the flat position, both wings immediately
begin to raise simultaneously.
Left wing stops at 12° and the right wing stops at 40°.
5 seconds after the right wing has reached 40° the left wing
lowers to 0°.
This position A is maintained for 30 minutes.
Simultaneously the left wing begins to raise and the right
wing begins to lower.
The right wing stops at 12° and the left wing continue to
raise up to 40°.
5 seconds after the left wing has reached 40° the right wing
lowers to 0°.
This position B is kept for 30 minutes.
The cycle will continue shifting between position A and B, until
the program is stopped.

4.7.2 Pause automatic operation

It is possible to pause a automatic operation and use
the hand control as in manual mode.
12
LED
Note
AUTO LED
active
P1 LED active
P2 LED active
The display
changes between
the selected
program and the
time to the next
reposition.
-P1/01 = program
1 and 01 minute
before next repo-
sition.
-P2/05= program
2 and 05 minutes
before next repo-
sition.
1.
Press and hold for 3 sec. R button, while an automatic pro-
gram is running, the wings will move to the flat position. Now
the manual mode can be used.
2.
The activated program before the pause action, will proceed
after 30 minutes of inactivity, 10 sec before the program starts
a BEEP is heard, if the inactivity period should be extended
press any button on the hand control after the BEEP.
3.
If the automatic program should start again right away, press
and hold R for sec 3 to activate.

4.8. Charging the battery

The battery is charging when the system is connected to the
power supply and turned on.
Charging of the battery is indicated by a dot in the bottom right
corner of the display.
The battery will be fully charged after 12 hours.
The battery is NOT charging when error E03 or E04
occurs.
4.9

Emergency release of the wing

In case of a power or motor failure, an emergency of the wing could
be necessary.
CAUTION !
Risk of injury
- A minimum of two persons is required for an emergency
release of the wing.
- When releasing the wing it might lower fast. Keep clear
the area under the wing and arms.
IMPORTANT!
- Before an emergency release of the wing, re-
move the plug from the mains socket.
1.
Both persons hold the wing.
2.
One of them locates the motor pin in question and pulls out
the safety pin.
3.
After the pin is removed, the motor will slide out of the sup-
port and slowly lower the wing.
IVCMAMED221119