Operation; Led Indicator; Checking The Glycol Sample Bottle - Thermo Scientific TSX Series Installation And Operation Manual

5.5 cu.ft. undercounter refrigerator
Table of Contents

Advertisement

6 Operation

6.1 LED Indicator

6.2 Checking the Glycol
Sample Bottle
Thermo Scientific TSX 5.5 cu.ft Undercounter Refrigerator
When you turn on your refrigerator, verify that the LED indicator is working.
The LED indicator appears on the front control panel next to the upper corner
of the display. The LED indicator illuminates green when the unit is on and
running. It illuminates red if there is a data log concern such as:
Low Battery
Full Memory
Data Logging Error
A power outage would result in a flashing red or green light every 15 seconds.
Flashing green indicates continuous data logging. Flashing red indicates a data
logging error. See the troubleshooting guide for ways to resolve the issue.
The glycol sample bottle is pre-installed inside your unit. The sample bottle
allows you to monitor the internal temperature of items being stored in the
refrigerator.
Verify the following:
1. The cord for the sample bottle is fully plugged into the back wall of the
interior of your unit.
2. The sample bottle is securely mounted in the clip provided.
3. The level of the propylene glycol in the bottle is full.
In the event that the bottle has broken during shipping, contact our service
department for a replacement bottle.