Installation; Leveling; Intended Use; Providing Ventilation - Thermo Scientific TSX Series Installation And Operation Manual

5.5 cu.ft. undercounter refrigerator
Table of Contents

Advertisement

Installation

5 Installation

5.1 Leveling

5.2 Intended Use

5.3 Providing
Ventilation
8
Installation and Operation
Adjust the leveling feet located underneath the refrigerator by turning clockwise
to extend the feet, counter clockwise to shorten them. It is important that the
unit is level.
Do not locate or store your refrigerator outdoors. This model was designed for
indoor use only and violation of this will void the terms of your warranty.
The refrigerators described in this manual are high performance units for
professional use. These products are intended for use as cold storage in research
use and as a general purpose laboratory refrigerator, storing samples or inventory
at operating temperatures between 3°C and 7°C.
It is not considered a medical device and has therefore not been registered with a
medical device regulatory body (e.g. FDA): that is, it has not been evaluated for
the storage of samples for diagnostic use or for samples to be re-introduced into
the body.
This unit is not intended for use in classified hazardous locations, nor to be used
for the storage of flammable or corrosive inventory.
Your TSX undercounter refrigerator is equipped with a ventilation area just above
the door handle. Do not block or cover the venting area as this can impact the
performance of the refrigerator.
The fans on your product will run at intervals of variable speeds and may turn off
completely. They do not run constantly. If the fans run frequently, the product
may require more space for ventilation. Please refer to the section on how to
clean and troubleshoot the fan for more information.