Service

Replacing the K-730 Filter

Remove all samples from the chambers before
replacing the filters.
1.
To access to the filter compartment, push
the safety lock lever (1) and lift the upper
part of the incubator containing the
chambers.
2.
Fasten the connectors to the inlet and
outlet of the K-730 Filter (2).
3.
Ensure the direction of flow shown on the
K-730 Filter label (3) is consistent with the
flow direction (4) shown inside the filter
compartment.
4.
Place the K-730 Filter in the filter holder
with the label facing up (5).
5.
Lower the upper part of the incubator.
CAUTION: Do not use the unit without a genuine K-Systems K-730 Filter.
CAUTION: Insert filter with the filter label facing up. If the filter is not placed correctly, it
can cause excessive gas consumption.

Disposal of the K-730 Filter

CAUTION: Contamination Hazard
As the filter may have been used for processing and treating infectious substances, it
might be contaminated. The used K-730 Filter should be placed in a sealed plastic bag
and labeled as biohazard material, then disposed of according to local requirements.
13
46
1
2
4
5
3