Section 13 - Service
WARNING: DO NOT disassemble or modify any part of the G210 InviCell
For the reliable and safe operation of this incubator it is strongly recommended that inspections and
services are performed as stated in the Service Plan below. Failing to follow this plan may cause the
unit to stop performing as intended and cause damage to embryos, blastocysts etc kept inside the
incubator.
Service Plan
Conducted by
Component name
Replace K-730 Filter Capsule
Replace UV light bulb
Replace HEPA Inline Filter for CO
Replace HEPA Inline Filter for N
Replace O
sensor*
2
Temperature and gas calibration
Replace Pump
Replace CO
sensor*
2
*Gas calibration should be performed after replacing O
On-screen Prompts
2
User
Every 3
months
X
gas
2
gas
2
2
The service symbol (1) appears in the main
screen when it's time for a service.
1
The Service tab shows when the unit's individual
parts need to be serviced (2).
This screen (3) also displays hardware and
software versions currently installed and the
unit's serial number.
3
Authorised Service Representative
Every
Every 3
year
years
X
X
X
X
X
X
and CO
sensors.
2
Service
Every 6
years
X
13
45