9. Thread the nut of the clear 15" Teflon tubing into
the top left port of the AutoInjector. Thread the
nut side of the black 72" tubing into the top right
port of the AutoInjector (Figure 4).
Left port
tubing is
connected to #3
position on the
injector.
Figure 4: Tubing location on the AutoInjector
Module
10. Route the clear 15" Teflon tubing behind the sam-
ple loop, then connect it to port #3 of the Injec-
tion valve and tighten the metal fitting with a ¼"
wrench (Figure 5).
#3 position
Figure 5: Tubing location
11. Take the open end of the black tubing connected
in Step 10 and route to a waste container or tee
into the existing waste lines.
12. The AutoInjector sample probe should be
connected to Port #2 of the injection valve if the
AutoInjector is being used without an
AutoSampler. If an AutoSampler is being used,
then the AutoSampler probe should be
connected to Port #2 instead.
Teledyne Isco
P.O. Box 82531, Lincoln, Nebraska, 68501 USA
Toll-free: (800) 775-2965 • Phone: (402) 464-0231 • Fax: (402) 465-3001
E-mail: IscoService@teledyne.com
Teledyne Isco is continually improving its products and reserves the right to change product
specifications, replacement parts, schematics, and instructions without notice.
13. Reconnect the power cord in the back of the unit
14. To verify that the system recognizes the pres-
Right port
tubing is
connected to
waste container.
Figure 6: Multiple Injection Run screen
Route
tubing
behind
loop
and power the ACCQPrep O
ence of the AutoInjector, press the play button
and the M
I
ULTIPLE
NJECTION
appear (Figure 6).
.
N
R
screen will
UN
Revised March 2018

Advertisement

loading