10 Notes on EMC (Electromagnetic Compatibility)
RF emissions
CISPR11
Harmonic emissions IEC
61000-3-2
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity for all equipment and systems
ICS AirCal is intended for use in the electromagnetic environment specified below. The user of ICS AirCal should ensure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short inter-
ruptions and voltage vari-
ations on power supply input
lines
IEC 61000-4-11
Voltage interruptions on
power supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
U
is the AC mains voltage prior to application of the test level.
T
Guidance and manufacturer's declaration - electromagnetic immunity - for equipment and systems within Professional Healthcare use environment
ICS AirCal is intended for use in the electromagnetic environment specified below. The user of ICS AirCal should ensure that it is used in such an environment.
28
Class A
ICS AirCal is suitable for use in hospital and clinic environments.
Note: The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class
Complies
B is normally required) this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as relocating or
re-orienting the equipment.
Complies
IEC 60601
Compliance level
test level
+/- 8 kV contact
+/- 8 kV contact
+/- 2 kV, +/- 4 kV,
+/- 2 kV, +/- 4 kV,
+/- 8 kV, +/- 15 kV air
+/- 8 kV, +/- 15 kV air
+/- 2 kV for power supply lines
+/- 2 kV for power supply lines
+/- 1 kV for input/output lines
+/- 1 kV for input/output lines
+/- 1 kV line(s) to line(s)
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
+/- 2 kV line(s) to earth
+/- 2 kV I/O line(s) to earth
+/- 2 kV I/O line(s) to earth
0 % U
; 0.5 cycle
0 % U
T
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
and 315°
and 315°
0 % U
; 1 cycle
0 % U
T
and
and
70 % U
; 25/30 cycles
70 % U
T
Single phase: at 0°
Single phase: at 0°
0 % U
; 250/300 cycles
0 % U
T
30 A/m
No relevant ports that could be
affected
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humid-
ity should be at least 30 %.
Mains power quality should be that of a typical com-
mercial or hospital environment.
Mains power quality should be that of a typical com-
mercial or hospital environment.
; 0.5 cycle
Mains power quality should be that of a typical com-
T
mercial or hospital environment. If the user of the ICS
AirCal requires continued operation during power
; 1 cycle
mains interruptions, it is recommended that the ICS
T
AirCal be powered from an uninterruptible power sup-
; 25/30 cycles
ply or a battery.
T
; 250/300 cycles
T
Power frequency magnetic fields should be at levels char-
acteristic of a typical location in a typical commercial or
hospital environment.
Otometrics - ICS AirCal