9 Technical Specifications
Standards
Safety
EMC
Performance
Stimulus temperature range
Cool stimulus
Warm stimulus
Temperature readout
Time to temperature
Temperature accuracy
Time range
Flow rate range
Flow rate accuracy
Dimensions
Width
Depth
Height
Weight
With empty reservoir
With full reservoir
Power supply
Input voltage
Power consumption
Fuse
Applied Part
The applied part of the ICS AirCal is the irrigator head.
26
Class I, Type BF, IPX0.
Complies with:
IEC 60601-1:2005+A1:2012
IEC 60601-1-6:2010+AMD1:2013
ANSI/AAMI ES60601-1:2005 + A1:2012
CAN/CSA-C22.2 NO. 60601-1:14
CAN/CSA-C22.2 No. 60601-1-6:11(R16)
IEC 60601-1-2
12° to 50°C
12° to 37°C (Recommended setting: 24°C)
37° to 50°C (Recommended setting: 50°C)
Digital
< 60 seconds
+/- 0.4°C
+/- 0.4°C 1 to 99 seconds (Recommended setting: 60 seconds)
4 - 10 liters per minute (Recommended setting: 8 liters per minute)
+/- 0.4 liters per minute
34.3 cm (13.5 inches)
30.5 cm (12 inches)
10.2 cm (4 inches)
7.9 kg (17.5 lbs)
8.2 kg (18 lbs)
100-240 VAC 50/60 Hz
<192 VA
T 1.6H 250V
Otometrics - ICS AirCal
Need help?
Do you have a question about the ICS AirCal and is the answer not in the manual?
Questions and answers