9. If the temperature does not reach 12°C in approximately 3 minutes or, if the over temperature alarm sounds, skip to
the procedure
10. Turn off the irrigator and unplug from the electrical wall outlet.
11. Remove the internal cover.
12. Check if the water reservoir is full. If not, place the funnel in the refill hole of the reservoir and add additional water.
13. Remove the funnel and replace the internal cover. Tighten the screws to secure.
14. Replace the external cover and tighten screws to secure.
15. Plug in the irrigator into the electrical wall outlet and turn on.
Note
The default airflow rate for caloric irrigation is 8 liters per minute. Shipping may cause the airflow knob to
move. You may have to reset the airflow.
Priming with bent tube
1. Take the syringe with bent tube and remove enough water from the reservoir to fill the syringe.
2. Put the end of the bent tube into the pump inlet hole. The tube will be facing toward the front of the irrigator.
3. While pressing the priming switch
4. Release the priming switch.
5. Press the priming switch to turn on and hold 5 seconds.
6. Release the priming switch to turn off and wait 5 seconds.
7. Repeat steps
Otometrics - ICS AirCal
Priming with bent tube
, inject the water into the inlet.
5
and
6
for a total of 10 times on and 10 times off.
13.
2 Getting started