INITIAL INSPECTION Carefully unpack the unit and accessories. Inspect for damages made in shipment. If any damage is found, notify your JENCO representative immediately. All packing materials should be saved until satisfactory operation is confirmed.
INSTALLING THE BATTERIES To insert the batteries into the meter, follow the procedure outlined below. Use a Philip screw driver and unlock the battery cover by turning the screw driver in the counter clockwise direction. After unlocking the screw, take off the battery cover (Fig.1) Replace the old batteries with new ones and install them on the correct polarity position.
B. Keys On/Off- Press and hold this key for 2 seconds to power on and shut off the meter. Cal- During normal operation, this key will change the lear mode from “Measure” mode to “Calibration” mode. Mode / Clear- In measure mode, press this key for 0 second to sequentially display Conductivity, TDS and Salinity.
OPERATIONAL PROCEDURES A. Calibration Calibration setup contains six sections: TDS Constant, Temperature Coefficient, Temperature Reference, Probe Basic Cell Constant, Standard Solutions Calibration and K Value Input. To access these sections: Connect the conductivity probe to the unit and turn the unit In conductivity, TDS or salinity mode, allow temperature reading to stabilize, press “Cal”...
Page 8
Probe Basic Cell Constant The main display shows the Cell Constant of the conductivity probe (calibrated previously or default, the deviation range is 70% ~ 130%). The secondary display shows the current selected cell constant which is either 0.01, 0.10, 1.00 or 10.0. Press the “Up”...
B. Conductivity Measurements Turn the unit on. Place the probe in the solution to be measured. Immerse the probe (at least 2” to 3” or 5~7cm from the tip). Shake the probe lightly to remove any trapped air bubbles in the conductivity cell. Press “Mode”...
"over " during measurements 1. Sample 1. Increase temperature sample temperature. undr < 0.0 ℃ 2. Replace cell. 2. Defective conductivity cell. [Note: If the meter still does not perform normally after the above measures are taken, call JENCO representative.]...