Download Print this page
Oakton pHTestr 30 Operating Instructions

Oakton pHTestr 30 Operating Instructions

Pocket tester
Hide thumbs Also See for pHTestr 30:

Advertisement

Quick Links

Operating Instructions
pHTestr
30 Pocket Tester
®
Large Screen
Waterproof pH/Temperature Tester
Double-Junction Sensor

Applications

• Aquaculture
• Food sectors
• Aquariums
• Hydroponices
• Boilers
• Labs
• Car washes
• Pools
• Ecology studies
• Sanitation plants

Before You Begin

Your instrument has been factory calibrated and usually works well out of
the box. However, after extended periods of non-use, it's best to soak the
sensor in electrode storage solution, pH buffer for 10 minutes or so. A brief
rinse with deionized (DI) water is OK, but avoid soaking or storing in deionized
water as this will shorten the pH electrode life. Prior to taking measurements,
periodic calibration with certified standards is recommended for best accuracy.

pH Buffer Set Selection

pHTestr features USA (pH 4.01, pH 7.00 and pH 10.01) or NIST (pH 4.01,
pH 6.86, and pH 9.18) standards. Select either one to suit your requirements.
1. While pressing the HOLD/ENT button, switch on the unit by pressing the
ON/OFF button.
2. Release the HOLD/ENT button. The display will flash either USA or NIST.
3. Press CAL button to toggle between the two buffer set standards.
4. Press the HOLD/ENT button to confirm the selection of the buffer set.
With meter
powered off, press
HOLD/ENT and
ON/OFF at the
same time. First
release ON/OFF
button, then
HOLD/ENT button.
Figure 1: Buffer Selection Sequence
1065O100_MAN_35634-30.indd 1

pH Calibration

Calibration should be done regularly, preferably once a week. You can calibrate
up to three points using either the USA or the NIST buffer set standards.
1. Press ON/OFF button to switch unit on.
2. Dip electrode about 2 to 3 cm into the pH standard buffer solution.
3. Press the CAL button to enter calibration mode. The 'CAL' indicator will be
shown. The upper display will show the measured reading based on the last
calibration while the lower display will indicate the pH standard buffer solution.
Note: To abort calibration, press the CAL button.
4. Allow about 2 minutes for the reading to stabilize before pressing the
HOLD/ENT button to confirm the first calibration point. The upper display
will be calibrated to the pH standard buffer solution and the lower display
will then search for the next pH standard buffer solution.
• Spas
5. Repeat with other buffers if necessary. Rinse electrode before dipping into
• Steam generators
next buffer.
• Water & wastewater
Note: The calibration mode allows you to perform up to three calibration points
treatment
before returning to the measurement mode automatically. However, if you opted
• Water quality testing
to have only one or two calibration points, simply skip the remaining calibration
points by exiting to the measurement mode by pressing the CAL button.
• More!
Figure 2: Example of pH Calibration Sequence
pH Measurement
1. Press the ON/OFF button to switch the unit on.
After buffer
selection, whole
2. Dip the electrode about 2 to 3 cm into the test solution. Stir and let the
LCD segment
reading stabilize. For semi-solid samples using pHSpear, rotation is ideal
lights up and
for sample contact.
goes back to
measurement
mode.
CAUTION: Testing dry samples is not accurate and can lead to sensor damage
or breakage. Soils must be wet and free of particulates that may scratch the
glass sensor. Excessive force into dry samples can cause glass breakage.
3. Note the pH value or press HOLD/ENT button to freeze the reading. To
release the reading, press HOLD/ENT again.
4. Press ON/OFF to turn off unit. If you do not press a button for 8.5 minutes,
the unit will automatically shut off to conserve batteries.
HOLD Function
This feature lets you freeze the display for a delayed observation.
1. Press HOLD/ENT button to freeze the measurement. A 'HOLD' indicator will
be displayed and the measurement will be frozen.
2. Press HOLD/ENT again to release the measurement. The 'HOLD' indicator
will not be displayed anymore indicating the held measurement is released.
User Reset
You can reset the pH calibration to the factory default by using the user reset
function. Buffer set selection and temperature user calibration are not affected
by the user reset function.
1. Switch off the unit.
2. While pressing the CAL button, press and release the ON/OFF button to
To do a 1 point calibration
enter the 'User Reset' selection menu. The screen will display 'rSt' on the
only, press CAL button at this
bottom display with a flashing 'nO' selection.
point to exit to the measurement
mode. Otherwise, proceed to
3. Use the CAL button to toggle between 'nO' and 'YES' selection.
second buffer for a second
buffer for a second point
• nO deactivates reset selection
calibration.
• YES activates the reset selection
To do a 2 point calibration only, press
4. Press the HOLD/ENT button to confirm the selection made.
CAL button at this point to exit to the
measurement mode. Otherwise,
5. If you have selected 'YES', the unit will show 'CO' momentarily and proceed
proceed to third buffer for a third
to the measurement mode with the calibration reset back to factory default
buffer for a third point calibration.
value.
After the third point calibration, the meter will
6. If 'nO' is selected, the unit will proceed to the measurement mode without
automatically return to the measurement mode.
any calibration reset performed.
At any point, an error message 'Er. 1' will be
displayed momentarily if the confirmed pH value
is not within the pH calibration window.
With meter
powered off, press
CAL and ON/OFF at
the same time. First
release ON/OFF
button, then
CAL button.
ATC - Automatic Temperature Compensation
Through its built-in temperature sensor, the measurement error due to the
changes in electrode sensitivity due to changes in temperature is compensated
to give the actual pH reading of the sample measured.
Note: To exit this program without confirming the calibration, press the
CAL button before the automatic confirmation takes place.
Figure 4: Example of HOLD Function
After reset, whole
LCD segment lits
up and goes back to
measurement mode.
Figure 5: User Reset Sequence
8/25/2017 7:15:07 AM

Advertisement

loading

Summary of Contents for Oakton pHTestr 30

  • Page 1: Operating Instructions

    Operating Instructions pH Calibration HOLD Function Calibration should be done regularly, preferably once a week. You can calibrate This feature lets you freeze the display for a delayed observation. up to three points using either the USA or the NIST buffer set standards. 1.
  • Page 2 6. Once the new temperature setting is reached, the new value is automatically Temperature calibration error of attempting to calibrate Er.0 Replacement sensor for pHTestr 30 35624-38 confirmed and returns to the measurement mode if no button is pressed after...