Model 35628-00 Microprocessor Based Pocket Size Temperature Tester
BEFORE FIRST USE: Remove plastic strips between batteries and contacts if present.
Caution: do not bend probe or expose the plastic probe connector to temperatures above
176°F (80°C).
CONNECTING THE PROBE: Remove the tester cap. Insert the probe plug into the jack at the
bottom of the TempTestr.
CALIBRATION: The TempTestr is factory calibrated to ensure accuracy within stated tolerances. It
is not necessary to calibrate unless it is desirable to achieve exact readout agreement with a work
standard at a specific temperature.
Select a reference thermometer such as an
temperature controlled water bath up to the calibration temperature point and note the reading.
Insert the TempTestr probe tip 3 inches into the water bath.
Press the TempTestr
again and let up both buttons. A "CA" will appear and the temperature will flash. Adjust the reading
±5° by pressing the
to confirm calibration. A "CO" will be displayed, then the calibrated temperature will be displayed
and the
function is cancelled. This calibration is retained until batteries are replaced at which
HOLD
time the TempTestr defaults to the factory calibration.
TEMPERATURE TESTING:
• Connect the probe and press
• Insert the probe tip where measurement is needed. Do not immerse probe above
plastic connector!
• Press "
" button if you wish to hold the reading ("
HOLD
Press it again to release it ("
• Press the on/off button to shut off.
• There is an automatic shut-off after 8.5 minutes to conserve batteries.
CHANGING BATTERIES: Flip up the battery compartment lid. Remove old batteries and replace
with fresh ones noting polarity as shown in the battery compartment.
SPECIFICATIONS:
Range:
Resolution:
Accuracy:
Operating Temp.: 32 to 122°F, 0 to 50°C
Battery/Life:
Size (meter only): 5.94"L x 1.65"W x 0.94"H (151 x 42 x 19 mm)
Probe Size:
Weight:
ERROR MESSAGES:
ER 1
means the batteries are low and should be replaced.
OR
means the temperature is out of range or the probe is not fully inserted into the jack
or is shorted.
NOTE: the probe is NOT interchangeable with other TempTestr probes and is NOT replaceable.
TempTestr
button, then press and hold the °F/°C button. Press the
HOLD
button until the desired temperature is displayed. Press the °F/°C button
HOLD
/
ON
OFF
" will be displayed momentarily).
HC
-20 to 220°F, -28 to 105°C
1°F, 1°C
±1°F from 20 to 160°F, 1%F.S. beyond. ±1°C through the entire range
3pcs. 1.4V Eveready EP675E / 70 hrs (Silver Oxide battery may be
substituted; use model 303. Typical life=50 hrs.)
Stem=4.5"L (115 mm) Overall=6.6"L (168 mm) Stem Dia.=0.125"(3.2 mm)
3.17 oz. (90 gm) (meter only) 0.25 oz. (7 gm) (probe only)
or
certified thermometer. Insert into a
NIST
ASTM