Download Print this page
Oakton pHTestr 30 Operating Manual

Oakton pHTestr 30 Operating Manual

Pocket testers
Hide thumbs Also See for pHTestr 30:

Advertisement

Quick Links

Operating Manual
pHTestr
30 Pocket Testers
®
Thank you for purchasing an Oakton
pHTestr
®
Please carefully read this operating manual before using the
product to obtain an accurate and reliable test result and avoid
unnecessary damage to the tester or sensor/probe.
Features
• Large easy-to-read LCD with
• No air bubbles generating in
3-color backlight
blue gel inner solution
• Accuracy of ±0.01 pH ± 1 digit
• Bulb protector can avoid
• 1, 2, or 3 calibration points
impacting. Remove it to rinse
• Auto buffer recognition: US or
the probe.
NIST
• Stability and battery status
Bulb
protector
icons
• Replaceable double-junction
sensor saves your money
• Auto power-off function
conserves battery life
• IP67 waterproof rating
Keypad Functions
Short press = <2 seconds Long press = >2 seconds
1. Short press to turn on the tester and long press to turn
off the tester.
2. When turned off, long press to enter parameter setting
MEAS
mode.
3. In measurement mode, short press to turn on backlight.
1. In measurement mode, short press to switch
parameter from pH to ORP (Oxidation Reduction
Potential); ORP probe sold separately.
2. In parameter setting mode, short press to change
parameter (unidirectional).
1. Long press to enter calibration mode.
2. In calibration mode, short press to confirm calibration.
CAL
3. When measured value is locked (HOLD icon), short
<
press to unlock.
Preparation Before Use
If it is first-time use or the tester hasn't been used for a long
time,pour some 3M KCL solution to the Fill line in the probe cap and
soak the probe for about 15 to 30 minutes. Users can store the
probe in the 3M KCL solution in the probe cap when the tester is not in
use to maintain the sensor's accuracy. To achieve maximum
accuracy, we recommend soaking the probe in the 3M KCL storage
solution overnight (12 hours) to activate the glassmembrane thoroughly.
Storing the pH30 tester dry will NOT cause any permanent damage.
It will only temporarily cause the probe to lose its sensitivity, which
can always be restored by soaking in the storage solution.
pH Calibration
1. Short press
2. Rinse the probe in distilled water and use tissue paper to
gentlydab off excess water.
3. Pour certain amount (about half volume of the calibration bottles)
of pH 7.00 pH and pH 4.00 buffer solutions in the corresponding
calibration bottles.
Pocket pH Tester.
®
4. Long press CAL/ key to enter calibration mode, short press
key to return to measurement mode.
5. Dip the probe in pH 7.00 buffer solution, stir gently, and allow it to
stand still in the buffer solution until a
stable reading is reached. When
measurement stability icon (  ) appears
and remains on the screen (see Fig 1);
then short press CAL/ key to complete
the one-point calibration. The tester
returns to measurement mode and
indication icon "M" will appear at the
bottom left of the screen, indicating the
middle point of calibration has been
completed.
6. For 2-point calibration, start with pH 7.00
It's detachable
buffer solution; once confirmed then rinse the probe in distilled
to rinse the
water and dry it with tissue paper. Dip the probe into pH 4.00
probe.
buffer solution, follow the steps above to complete 2-point
calibration. Indication icons "L" and "M" will appear on the bottom
left of the screen.
7. If necessary to perform a 3-point calibration, rinse probe with
water, dip the probe into pH 10.01 buffer solution (sold
separately), and follow the steps above. Indication icons "L", "M",
and "H" will appear on the bottom left of the screen.
Notes
A. Tester can perform 1 to 3 points automatic calibration. Please
note that pH 7.00 (USA Standard) or pH 6.86 buffer solution
(NIST Standard) must be used to conduct 1st point calibration.
Then use other buffer solution to conduct 2nd or 3rd point
calibration. Tester will automatically recognize five kinds of pH
buffer solutions. For details, please refer to the following table:
Calibration
USA series
1-point
7.00 pH
7.00 pH,
4.00 pH or
1.68 pH
2-point
7.00 pH,
10.01 pH or
12.45 pH
7.00 pH,
4.00 or
3-point
1.68 pH,
10.01 or
12.45 pH
B. The buffer solutions poured into the calibration bottles are NOT
for one-time use. They can be used multiple times as long as
they are not contaminated, and the bottles are covered when not
in use. After that, we recommend replacing the buffer solutions in
the calibration bottles with new ones that are in the buffer bottles
(50 mL) to keep the accuracy of the standard buffer solutions. Do
not pour used buffer solutions back into the buffer bottles in case
of contamination.
/MEAS key to turn on.
Indication
NIST series
Recommended
icons
6.86 pH
M
Accuracy:≥0.1 pH
6.86 pH,
L M
4.01 pH or
Range:<7.00 pH
1.68 pH
6.86 pH,
9.18 pH or
M H
Range:<7.00 pH
12.45 pH
6.86 pH,
4.01 or
Range:
1.68 pH,
L M H
0 to 14.00 pH
9.18 or
12.45 pH
C. The tester has self-diagnostic functions:
Self-diagnostic
Symbol
information
Wrong pH buffer
solution or the range
of calibrationsolution
/MEAS
exceeds standard.
Press CAL/ key
when reading is not
stable during
measurement.
* The 1st point calibration must be 7.00 pH. Perform the 2nd point
calibration (4.00 pH) immediately after the 1st point. Do NOT turn
off the meter before you conduct 2nd point calibration. If the meter
is turned off after 1st point calibration, user will need to restart the
calibration process with the 7.00 pH first and the 4.00 pH following
after. Calibrating directly in pH 4.00 after turning meter off and
back on will cause "Er1".
pH Measurement
1. Short press
/MEAS key to turn on the tester. Rinse the probe in
distilled water, and gently dab off excess water with tissue paper
(do not rub or wipe the glass sensor).
2. Dip the probe in sample solution, stir gently, and allow it to stand
until a stable reading is reached. Take readings after stability
indicator icon (  ) appears and stays on the screen.
Notes
A. Applications:
Model/Probe
Regular water solutions' pH measurement
pH30/Bulb probe
such as hydroponics, aquaculture, pools and
spas, water treatment, brewing, etc.
Special Notes
• The pH probe must be rinsed thoroughly after each use. Soap
water should be used to clean off any grease or other
contaminants.
• These pH testers will NOT give accurate or stable pH readings
when testing distilled or deionized water. This is because distilled
or deionized water do not have enough ions present for the
electrode to function properly. To measure distilled or deionized
water's pH, users need to use a specialized instrument technique.
When testing purified water like spring water or drinking water, it
will take longer for the readings to get stabilized (typically 3 to 5
minutes) because there are very few ions left to be detected by the
sensor in those purified water.
• Do NOT store probe in distilled or deionized purified water
because that will cause permanent damage to the pH probe.
Purified water is only recommended for rinsing the probe. The
probe should be stored in 3M KCL pH electrode storage solution
(SKU 00653-04) for best accuracy.
Setting the Parameters
When tester is turned off, long press
parameter setting. Short press MODE/△ key to switch from P1 to
Checking and how to fix
1. Check whether pH buffer solution
is correct (1st point calibration
must be 7.00).
2. Check whether the probe is
damaged.
3. Check if there is any air bubble in
the glass bulb sensor.
Wait for the measurement
stability icon (  ) to appear and
stay, then press CAL/ key.
Application
/MEAS key to enter

Advertisement

loading

Summary of Contents for Oakton pHTestr 30

  • Page 1 7.00 pH and pH 4.00 buffer solutions in the corresponding must be 7.00). solution or the range calibration bottles. 2. Check whether the probe is of calibrationsolution Thank you for purchasing an Oakton pHTestr Pocket pH Tester. ® ® damaged.
  • Page 2 — 35634-68 (double junction) Oakton Instruments reserves the right to update the information in Select “Yes” to recover instrument calibration to the theoretical value this manual without giving notice in advance. (pH value in zero potential is 7.00 pH, slope is 100%), parameter Probe Replacement setting returns to initial value.