Backlight Brightness Setting; Power-On Parameter Setting - East Tester ET4501 User Manual

Lcr benchtop digital bridge
Table of Contents

Advertisement

ET4501/4502/4510 台式 LCR 数字电桥操作手册
Step 2: Press the
identification of open circuit, short circuit, open circuit OPEN, short circuit SHORT) with the
interface as shown in Fig.4-18 (the number will be added 1 along with the correction progress,).
If the correction is successful, the word "success" is displayed in the position where "Digital" is
displayed. If it fails, "FAIL" will be displayed as shown in Fig.4-19; press the
again to return to Interface of Measurement Display.
Fig 4-18

4.2.18 Backlight Brightness Setting

Step 1: Turn on the instrument to enter the interface of measurement display, and press the
key to enter the Interface of System Setting.
Step 2: Move the cursor to "Brightness Adjustment" through the up and down direction keys,
and switch the backlight brightness (30%, 50%, 70%, 100%) through the left and right direction
keys.

4.2.19 Power-on Parameter Setting

Step 1: Turn on the instrument to enter the interface of measurement display, and press the
key to enter the Interface of System Setting。
Step 2: Move the cursor to the "Power-on Setting " via the up and down direction keys, and
then switch the "Power-on Setting " (default value, previous value) through the left and right
direction keys.
杭州中创电子有限公司
Fig 4-17
key to carry out open or short circuit correction (automatic
Fig 4-19
key
18