East Tester ET4501 User Manual page 24

Lcr benchtop digital bridge
Table of Contents

Advertisement

ET4501/4502/4510 台式 LCR 数字电桥操作手册
 Zero value adjustment : open circuit and short circuit zeroed before the test
 Warm-up time:> 30 minutes
Calibration time: 12 months
C: 0.20%*(1+ Cx/Cmax+ Cmin/Cx )(1+Dx)(1+ks+kv+kf);
L: 0.20%*(1+ Lx/Lmax+ Lmin/Lx )(1+1/Qx)(1+ks+kv+kf);
Z: 0.20%*(1+ Zx/Zmax+ Zmin/Zx )(1+ks+kv+kf);
R: 0.20%*(1+ Rx/Rmax+ Rmin/Rx )(1+Qx)(1+ks+kv+kf);
D: 0.20%*(1+ Zx/Zmax+ Zmin/Zx )(1+Dx+Dx*Dx)(1+ks+kv+kf );
Q: 0.20%*(1+ Zx/Zmax+ Zmin/Zx )(Qx+1/Qx )(1+ks+kv+kf );
Wherein
1. L, C, R, Z are relative errors; D, Q, θ are absolute errors
2. Where the subscript is x, it is the measurement of this parameter; where the subscript is max, it
is the maximum value and where the subscript is min, it is the minimum value
3. Ks is the speed factor, kv is the voltage factor, kf is the frequency factor
The maximum and minimum values of the measurement parameters that affect accuracy
Frequency(Hz
100
Cmax
800
Cmin
1500
Lmax
1590
Lmin
3.2
Z/Rmax
Z/Rmin
Frequency(Hz
15K
Cmax
5.3
Cmin
10
Lmax
10.6
Lmin
0.021
Z/Rmax
Z/Rmin
Wherein: the unit of Cmax is uF; the unit of Cmin is pF; the unit of Lmax is H; the unit of Lmin is
mH; the unit of Zmax / Rmax is MΩ; the unit of Zmin / Rmin is in Ω;
Measurement speed error factor ks:
Speed mode
slow
ks
0
Measurement speed error factor kf:
Frequency (Hz) 100~1K
kf
0
Measurement speed error factor kv:
level ( mV) 100
kv
10
杭州中创电子有限公司
120
200
400
667
400
200
1250
750
375
1325
795
397.5
2.6
1.6
0.8
20K
40K
4
2
7.5
3.75
7.95
3.98
0.016
0.008
medium
1
2K~10K
15K
0.5
300
600
1000
3
1
0
800
1K
2K
100
80
40
187
150
75
198.8
159
79.5
0.4
0.32
0.16
1
1.59
50K
1.6
3
3.18
0.0064
1
1.59
fast
8
20K
40K
50K
1
2
1500
2000
1
2
4K
8K
10K
20
10
8
37.5
18.7
15
39.8
19.9
15.9
0..08
0.04
0.032
80K
100K
1
0.8
1.87
1.5
1.99
1.59
0.004
0.0032
80K
100K
3
21

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Et4510Et4502Et45 series

Table of Contents