East Tester ET4501 User Manual page 7

Lcr benchtop digital bridge
Table of Contents

Advertisement

ET4501/4502/4510 台式 LCR 数字电桥操作手册
Measurement
0.0001Ω - 99.99MΩ
Range(R)
Basic Accuracy
0.2%
Test Frequency
10~10kHz adjust
continuously ,1Hz
(Hz)
stepping
Measurement
2 times/second (slow), 4 times/second (medium speed), 8 times/sec (fast)
Display Speed
Internal Bias
0-1500mV adjustable, 1mV stepping
Test Level
Correction
Open circuit correction, short circuit correction
Function
Screening
The screening limit can be set from -50% to + 50% with fixed points of
1%, 5%, 10% and 20%
function
Selection
of
Five gear sorting, third gear qualified, one gear failed, one ancillary gear
comparator
Communication
Standard configuration: USB, RS232 (or 485), Handle interface; Optional:
GPIB, USB Host
Interface
Support DCR, electrolytic capacitor measurement mode, adjustable
Others
backlight brightness, optional in English and Chinese
2.1
General Technical Specifications
●Power supply voltage: 220V AC ± 10%, or 110V AC ± 10%, 45-65Hz
●Working environment: 0 ~ 40 ℃, relative humidity <80%.
●Storage environment: -10 ~ 50 ℃, relative humidity <80%.
●Dimensions: 265mm * 105mm * 305mm (width * height* depth).
●Mass: 2.3kg.
●Power consumption: <10W
杭州中创电子有限公司
10~20kHzadjust
continuously ,1Hz
stepping
0.1~2V electrical level adjust continuously,1mVstepping
10~100kHzadjust
continuously ,1Hz
stepping
4

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Et4510Et4502Et45 series

Table of Contents