Metal components
Method: Wipe off with a wet cloth or soft brush. Water
may be pressurized, but not high pressure or steam.
Max. temperature: 40 °C
Solvent/chemicals: Household detergent or soap and water,
6-8 pH
Wood (including textile straps on side rails, if existing)
Method: Wipe off with a wet cloth or soft brush.
Max. temperature: 40 °C
Solvent/chemicals: Household detergent or soap and water,
6-8 pH
Textiles (including upholstery and mattresses)
See attached label on each product.
1579562-B

6.4 Lubrication

Lubrication plan
We recommend lubricating the bed according to the
following instructions:
Points of rotation in mattress support and base frame
- lubricate with medically clean oil.
Motor attachment points to mattress platform -
lubricate with medically clean oil.
All of the motors' tension rod - lubricate with medically
clean oil.
All motor bearings - lubricate with medically clean oil.
The wooden side rails gliding system must not be
lubricated with oil - otherwise the wooden bars will
move sluggishly.