Emergency Release Of A Mattress Support Section; Mattress Support Extension - Invacare Etude Plus HC User Manual

Bed
Hide thumbs Also See for Etude Plus HC:
Table of Contents

Advertisement

Invacare® Etude Plus® HC
1. Up: Lift the mattress handle on the leg section.
2. Down: Lift the mattress handle on the leg section right
up and then lower the leg section.
4.9 Emergency release of a mattress support
section
In case of a power- or motor failure, an emergency release
of the back, thigh or leg section could be necessary . An
emergency release of the height adjustment is NOT possible.
CAUTION!
Risk of injury
– A minimum of two persons is required for
an emergency release of a mattress support
section.
– When releasing a mattress support section,
it might lower fast. Do not reach under the
mattress support while lowering it.
IMPORTANT!
– Before an emergency release of the mattress
support, remove the plug from the mains
socket.
14
1. Both persons hold the mattress section.
2. One of them locates the motor in question and pulls
out the safety pin.
3. Both persons slowly lower the mattress section until it
is completely down.

4.10 Mattress support extension

For users, taller than two meters it is recommended
to use a mattress support extension.
Fitting the mattress support extension
1. Disassemble the bed and remove the inserts.
2. Mount the mattress support extension between both
parts of the mattress support and reassemble the bed.
1579562-B

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents