Cleaning And Disinfection; General Safety Information; Cleaning Intervals; Cleaning By Hand - Invacare Medley Ergo User Manual

Bed
Hide thumbs Also See for Medley Ergo:
Table of Contents

Advertisement

Invacare® Medley® Ergo
q
Check the function of the brakes.
q
Check the locking function(s) of the side rails.
q
Check that all bed slats are intact and not loose.

6.5 Cleaning and Disinfection

6.5.1 General safety information

CAUTION!
Risk of contamination
– Take precautions for yourself and use
appropriate protective equipment.
CAUTION!
Risk of electric shock and product damage
– Switch off the device and disconnect from
mains, if applicable.
– When cleaning electronic components consider
their protection class regarding water ingress.
– Make sure that no water splashes to the plug
or the wall outlet.
– Do not touch the power socket with wet hands.
NOTICE!
Wrong fluids or methods can harm or damage
the product.
– All cleaning agents and disinfectants used must
be effective, compatible with one another and
must protect the materials they are used to
clean.
– Never use corrosive fluids (alkalines, acid etc.)
or abrasive cleaning agents. We recommend
an ordinary household cleaning agent such as
dishwashing liquid, if not specified otherwise in
the cleaning instructions.
– Never use a solvent (cellulose thinner, acetone
etc.) that changes the structure of the plastic
or dissolves the attached labels.
– Always make sure that the product is completely
dried before taking into use again.
For cleaning and disinfection in clinical or long-term
care environments, follow your in-house procedures.

6.5.2 Cleaning Intervals

NOTICE!
Regular cleaning and disinfection enhance smooth
operation, increases the service life and prevents
contamination.
Clean and disinfect the product:
– regularly while in use,
– before and after any service procedure,
– when it has been in contact with any body
fluids,
– before using it for a new user.

6.5.3 Cleaning by Hand

All components (excluding removable textiles)
22
Cleaning agent: We recommend using a mild detergent
with neutral or near neutral pH (5-9).
Commercially available consumer products like
hand dishwashing liquid or all-purpose cleaners
can be used. Read the instructions on the
label of your cleaning agent and use it at the
indicated concentration.
Max. water temperature: 40 °C
1. Wipe surfaces thoroughly with a soft cloth slightly
moistened with detergent solution until all visible soil
is removed.
2. Remove all detergent residue with a clean damp cloth.
3. Wipe dry using a clean soft cloth.
Removable textiles (including upholstery and mattresses)
See attached label on each product.

6.5.4 Disinfection Instructions

In Domestic Care
Disinfectant: We recommend using an alcohol-based
surface disinfectant (with 70-90% alcohol).
Read the instructions on the label of your
disinfectant. It provides information on the
activity spectrum (bacteria, fungi and/or viruses),
material compatibility and the correct exposure
time.
1. Ensure the surfaces are cleaned before disinfection.
2. Moisten a soft cloth and wipe-disinfect all accessible
surfaces and keep them moistened for the exposure
time stated on the label of the disinfectant.
3. Allow the product to air-dry.
In Institutional Care
Follow your in-house disinfection procedures and only use
the disinfectants and methods specified therein.

6.6 Lubrication

We recommend lubricating the bed according to the
following instruction after service and cleaning or whenever
it is necessary to reduce friction.
Part of bed
Points of rotation in mattress
support
Motor attachment points on
mattress support
Motor bearings and tension
rods
The wooden side rails gliding system must not be
lubricated with oil - otherwise the wooden bars will
move sluggishly.
For ordering of correct oil, contact your Invacare
provider.
Lubrication method
Oil (medically clean)
Oil (medically clean)
Oil (medically clean)
60128481-C

Advertisement

Table of Contents
loading

This manual is also suitable for:

Medley ergo low

Table of Contents