Automated Cleaning And Disinfection - W&H proxeo TWIST PL-40 H Instructions For Use Manual

Hide thumbs Also See for proxeo TWIST PL-40 H:
Table of Contents

Advertisement

Hygiene and maintenance
Handpiece sleeve
W&H recommends automated cleaning and disinfection using a washer-disinfector (WD).
Read the notes, follow the instructions and heed the warnings provided by the manufacturers of washer-disinfectors,
cleaning agents and/or disinfectants.
> The handpiece drive, the handpiece holder and the foot control are not approved for automated processing in a
washer-disinfector and for sterilization.
Evidence of the medical device's basic suitability for effective automated disinfection was provided by an independent test
laboratory using the »Miele PG 8582 CD« washer-disinfector (Miele & Cie. KG, Gütersloh) and the »Dr. Weigert neodisher®
MediClean forte« cleaning agent (Dr. Weigert GmbH & Co. KG, Hamburg).
> Cleaning at 55°C (131°F) – 5 minutes
> Disinfection at 93°C (200°F) – 5 minutes

Automated cleaning and disinfection

33