Hygiene and maintenance
W&H recommends automated cleaning and disinfection using a washer-disinfector (WD).
> Read the notes, follow the instructions and heed the warnings provided by the manufacturers of
washer-disinfectors, cleaning agents and/or disinfectants.
Evidence of the medical device's basic suitability for effective automated disinfection was provided by an
independent test laboratory using the »Miele PG 8582 CD« washer-disinfector (Miele & Cie. KG, Gütersloh)
and the »Dr. Weigert neodisher® MediClean forte« cleaning agent (Dr. Weigert GmbH & Co. KG, Hamburg).
> Cleaning at 55°C (131°F) – 5 minutes
> Disinfection at 93°C (200°F) – 5 minutes
36