Hygiene and maintenance

Automated cleaning and disinfection

W&H recommends automated cleaning and disinfection using a washer-disinfector (WD).
> Read the notes, follow the instructions and heed the warnings provided by the manufacturers of
washer-disinfectors, cleaning agents and/or disinfectants.
Evidence of the medical device's basic suitability for effective automated disinfection was provided by an
independent test laboratory using the "Miele PG 8582 CD" washer disinfector (Miele & Cie. KG, Gütersloh) and
the "Dr. Weigert neodisher® MediClean forte" cleaning agent (Dr. Weigert GmbH & Co. KG, Hamburg) according
to ISO 15883.
> Cleaning at 55 °C (131 °F) – 5 minutes
> Disinfection at 93 °C (200 °F) – 5 minutes