Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections except the front and rear
CAUTION
panel bulkhead connectors. Torque these connections to 21 in-lb.
About Installing the Upgrade
Products affected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .N5241A and N5242A Option 200
Installation to be performed by . . . . . . . . . . . . . . . . . . . . . . . .Keysight service center or personnel qualified by
Estimated installation time . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 hours
Estimated adjustment time. . . . . . . . . . . . . . . . . . . . . . . . . . . .1 hour
Estimated full instrument calibration time . . . . . . . . . . . . . . .4.5 hours