Download Print this page

Keysight N5242-60115 Installation Note page 24

Add 4-port capability upgrade kit

Advertisement

60074) in the analyzer as shown. Secure both boards with two screws (item
in-lbs.
2. Connect cable W2 (N5222-20090) between the A8 26.5 GHz source 2 board and the A13 13.5 GHz source 2
synthesizer board as shown. Be sure to position the cable in the wire looms as shown. Tighten the cable
connectors to 10 in-lbs using a 5/16-in torque wrench.
3. Connect the loose end of new gray flex cable W22 (N5242-60030) on the A10 frequency reference board
(J7). (The other end of this cable was previously connected to J5 of the source 2 synthesizer board
Figure 9-2
Second Source Boards Installation
Installation Note N5242-90115
; 0515-0380) and torque to 21
.)
23

Advertisement