Initial Instrument Functionality Check

Power on the instrument and allow the instrument to boot up, run the alignments and display the
measurement screen. The instrument will probably display a spectrum analyzer screen and you will
see the instrument sweeping.
There should be no alignment failures. If there are failures, investigate and fix the problem before
continuing.
Instrument Software Installation
This upgrade requires instrument software version A.17.04 or later.
To verify the current version, press System, Show System and look for the "Instrument S/W
Revision".
Also look at the options section and assure N9040B-RT1 or RT2 is present, and also that
N9040B-B2X or B5X is present.
The latest revision of the X-Series Signal Analyzers software may be downloaded from:
http://www.keysight.com/find/xseries_software

Hardware Installation

1. Power down the instrument, wait till the standby light comes on, and remove the power cord.
2. Remove the instrument side strap handles, bottom and rear feet, outer case, and top brace.
3. Remove the eight screws holding each side of the front frame to the instrument chassis. A
T-20 and a T-10 Torx driver are required. Save screws for reuse. The front frame is being
loosened so the right side chassis can be removed in the next step.
4. Remove the right side chassis (RF Side Bracket) by removing all screws (see the UXA Service
Guide, Assembly Replacement section, A28 Wideband Extension Board Option RTS). On some
instruments there may be a small bracket that connects the right side chassis to the top brace.
Remove this bracket if present.
Very carefully pull the front frame forward a few millimeters (the external mixing cable and
ribbon cable are still attached!) so the pem in the right side chassis will clear the cutout in the
front frame casting. Pull the right side chassis up and out of the instrument.
5. This procedure assumes the Wideband IF boards (Option B2X or Option B5X) are already
installed in the instrument.