Keysight Technologies DAQ970A Programming Manual page 295

Data acquisition system
Hide thumbs Also See for DAQ970A:
Table of Contents

Advertisement

Standard Operation Register
Bit num-
Bit name
ber
0
Calibrating
1
Self Test
2
Digitizing
(for DAQM909A Digitizer module)
3
Waiting For Trigger
(for DAQM909A Digitizer module)
4
Scanning
5
Waiting For Trigger
(for Scan/DMM Digitize mode)
6
Not Used
7
USB MSD detected
8
Configuration Changed
9
Reading Memory Threshold
10
Instrument Locked
11
Settings Changed
12
Not Used
13
Global Error
14
Not Used
15
Not Used
Keysight DAQ970A/DAQ973A Programming Guide
Decimal value
Definition
1
The instrument is calibrating.
2
The instrument is doing a self-test.
4
One or more digitizer channels have been initiated for digitizing.
8
All initiated digitizer channels are in "wait-for-trigger" state.
16
The instrument is scanning.
32
The instrument is waiting for a trigger.
64
(Always returns 0)
128
A USB mass storage device (USB drive) has been detected.
256
The instrument configuration has changed via front panel since the
last INIT, READ? or MEASure?.
512
The number of readings in memory has exceeded the memory
threshold setting (
command)
1024
The instrument is locked (
2048
The instrument configuration has changed via front panel or SCPI
since the last INIT, READ? or MEASure?. Event only, condition
register returns 0.
4096
(Always returns 0)
8192
An error is in the global error queue.
16384
(Always returns 0)
32768
(Always returns 0)
2         SCPI Programming
DATA:POINts:EVENt:THReshold
SYSTem:LOCK
command)
295