*RST
Reset Command . Resets the instrument to the Factory configuration. See
of the instrument's Factory configuration.
Parameter
(none)
Resets the instrument: *RST
– This command does not affect any stored instrument states (*SAV), and non-volatile settings such as I/O con-
figuration, user preferences, or the calibration security state.
*SAV {0|1|2|3|4|5}
*RCL {0|1|2|3|4|5}
Save and recall command . Save the current instrument state to, or recalls the instrument state from, the state file
STATE_<n>.sta in the root folder of the internal flash file system, where <n> is the location number specified.
Parameter
{0|1|2|3|4|5}
Save the instrument state to location 1: *SAV 1
– *SAV < n > is equivalent to
– *RCL < n > is equivalent to
– LAN I/O configuration is not saved by a *SAV operation or recalled by a *RCL. Only instrument configuration is
recalled.
– State "0" is the state of the instrument at power down. The state file created by *SAV 0, called STATE_0.sta, is
overwritten with the power-down state of the instrument when power is cycled.
– The instrument state enable settings are saved in non-volatile memory. They do not change when power has
been off, after a Factory Reset (*RST), or after an Instrument Preset (SYSTem:PRESet).
Keysight DAQ970A/DAQ973A Programming Guide
Typical return
(none)
Typical return
(none)
MMEMory:STORe:STATe
MMEMory:LOAD:STATe
Factory Reset State
"INT:\STATE_ n .sta".
"INT:\STATE_ n .sta".
2 SCPI Programming
for a complete listing
151