[SENSe:]TOTalize:CLEar:IMMediate [(@<ch_list>)]
Clears the count on the specified totalizer channels immediately.
Parameter
(none)
Clears the count on totalizer channel immediately: TOT:CLEAR:IMM
[SENSe:]TOTalize:DATA? [(@<ch_list>)]
This query reads the count on the specified totalizer channels. If you have configured the count to be reset when it is
read (see
CONFigure:TOTalize
after it is read. The count is reset regardless of whether the specified channels are in a scan list or even whether a
scan is in progress.
Parameter
(none)
Reads the count on totalizer channel: TOT:DATA?
– The maximum count is 67,108,863 (2
– The output from this command is affected by the settings of the
the formats selected, each reading may or may not be stored with measurement units, time stamp, channel num-
ber, and alarm status information.
[SENSe:]TOTalize:SLOPe {POSitive | NEGative} [, (@<ch_list>)]
[SENSe:]TOTalize:SLOPe? [(@<ch_list>)]
Configures the specified totalizer channels to count on the rising edge (positive) or falling edge (negative) of the
input signal.
Parameter
{POSitive | NEGative}
Default: POS
Configures totalizer channel to count on the negative edge: TOT:SLOP NEG
– This parameter is set to its default value after a Factory Reset
Keysight DAQ970A/DAQ973A Programming Guide
Typical return
(none)
and
[SENSe:]TOTalize:TYPE
Typical return
1.32130000E+03
26
- 1). The count rolls over to 0 after reaching the maximum allowed value.
Typical return
POS or NEG
commands), then this command will reset the count to 0
FORMat:READing
(*RST
command).
2 SCPI Programming
commands. Depending on
281