Transcend 365 miniCPAP User Manual page 55

Hide thumbs Also See for 365 miniCPAP:
Table of Contents

Advertisement

Transcend 365 miniCPAP User Manual
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601 test
Compliance level
level
±8 kV contact
±2, 4, 8 kV contact ±2,
4, 8, 15 kV air
±2, 4, 8, 15 kV air
±2 kV for power
±2 kV for power supply
supply lines
lines
±1 kV for
±1 kV for signal
input/output lines
input/output lines
±1 kV differential
±0.5*, 1 kV differential
mode
mode
<5% U
(>95% dip
<5% U
T
in U
) for 0.5 cycle
U
T
T
40% U
(60% dip in
40% U
T
U
) for 1 cycle
for 1 cycle
T
70% U
(30% dip in
70% U
T
U
) for 25 cycles
for 25 cycles
T
<5% U
(>95% dip
<5% U
T
in U
for 5 sec)
for 5 sec)
T
30 A/m
30 A/m
3 Vrms
3 Vrms
150 kHz to 80 MHz
150 kHz to 80 MHz*
10 V/m
10 V/m
80 MHz to 2.7 GHz
80 MHz to 2.7 GHz
(>95% dip in
T
) for 0.5 cycle
(60% dip in U
)
T
T
(30% dip in U
)
T
T
(>95% dip in U
T
T
Page 55
Electromagnetic
environment—guidance
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Line power quality should be
that of a typical commercial or
hospital environment.
Line power quality should be
that of a typical commercial or
hospital environment.
Line power quality should be
that of a typical commercial or
hospital environment. If the user
of the Transcend 365 miniCPAP
requires continued operation
during power line interruptions,
it is recommended that the
Transcend 365 miniCPAP be
powered from the battery.
NOTE:
U
is the A.C. line
T
voltage before
application of the test
level.
Power frequency magnetic fields
should be at levels characteristic
of a typical commercial or
hospital environment.
Recommended separation
distance: d = 1.17
P
Recommended separation
distance: d = 0.35
80 MHz to
P
800MHz
Recommended separation
distance: d = 0.70
800MHz to
P
2.5 GHz
where
P
is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than the
compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
d