Caring For Your Transcend 365 Minicpap; Cleaning The Unit; Caring For The Humidifier Reservoir; Cleaning Your Mask And Air Hose - Transcend 365 miniCPAP User Manual

Hide thumbs Also See for 365 miniCPAP:
Table of Contents

Advertisement

Transcend 365 miniCPAP User Manual
Do not use harsh or abrasive cleaning agents to clean the device or any
components.
Do not attempt to sterilize the Transcend 365 miniCPAP.
Do not place cleaning materials, such as a cloth or liquid, into the device air inlet or
air outlet connector.

Cleaning the Unit

Follow these instructions to clean the exterior of the Transcend 365 miniCPAP:
• Wipe the device with a dry, lint-free cotton cloth
Exterior cleaning of the device should be performed as follows:
Product
Transcend 365
miniCPAP
Warning:
Do not submerge the Transcend 365 miniCPAP, power supply or power
accessory (battery) in liquid. Do not allow liquid or cleaning solution to enter the device.

Caring for the Humidifier Reservoir

Follow these instructions to care for the humidifier reservoir.
1.
Rinse the reservoir with distilled water weekly.
2.
Reattach the humidifier reservoir to the PAP device.
3.
Press and hold the Up and Down buttons for three seconds to start Drying Mode. This
mode will last 30 minutes unless you interrupt it by pressing any button. After the 30-
minute drying cycle, the blower will turn off and the device will automatically enter
Standby Mode. Dry mode flushes your CPAP device, tubing and mask of any residual
moisture.
4.
Refill the reservoir with water prior to next use.
NOTE:
The humidifier reservoir should be replaced every 6 months or when any
leaking or physical damage is noted.

Cleaning Your Mask and Air Hose

Follow the manufacturer's instructions on cleaning your mask and air hose, including
appropriate detergents and frequency of cleaning.
Periodic Cleaning Cycle
Weekly
Product Service Life
5-Year
Page 38

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents