General Commands - Hameg HM8131-2 Manual

Function generator
Hide thumbs Also See for HM8131-2:
Table of Contents

Advertisement

G e n e r a l c o m m a n d s

General commands

(memory card commands excepted)
Initialization
CLR
Initialization of the HM8131-2. Stored arbitrary
functions will remain untouched.
RST
Same as before except that any stored arbitrary
functions will be erased and replaced by those
from the factory.
General commands
STA?
Reading the instrument confi guration
VER?
Reading the software version
SNR?
Reading the serial number
ID?
Reading the instrument ID (short version)
*IDN?
Same, long version
TRG
Generation of a trigger pulse
STO:x
Store instrument status (0 .. 9)
RCL.x
Recall instrument status (0 .. 9)
Bus commands
LK0
Erase "local lockout"
LK1
Activate "local lockout"
RM0
Reset to manual operation
RM1
Set to remote control
OT0
Switch off output
OT1
Switch on output
RFX
Activate external reference frequency
RFI
Activate internal reference frequency
BPO
Turn off beep
BPS
Turn on beep medium
BPL
Turn on beep loud
Signal function
SIN
Sine
SQR
Square wave
TRI
Trinagle
RMP
Rising ramp
RMN
Falling ramp
ARB4
Arbitrary signal 4k
ARB16
Same 16k
PINK
Pink noise
NOISE
White noise
Signalparameter
FRQ?
Read frequency
FRQ.xxx
Programming a frequency
DFR
Switch display to frequency
VPP
Voltage display in Vpp
VRMS
Voltage display in Vrms
AMP7
Read actual amplitude
AMP.xxx
Programming an amplitude
DAM
Switch display to amplitude
OFS?
Read offset
OFS.xxx
Programming an offset
DOF
Display offset
PHA?
Read phase
PHA.xxx
Programming of a phase
Typs of running modes
CTM
Continuous mode = free running
TRM
Trigger mode
TRM+
Triggering on positive slope
TRM–
Same on negative slope
GTM
Gated mode
64
Subject to change without notice
GTM+
Gated mode on High level
GTM–
Same on Low level
TRP+
Triggering on High level
TRP–
Same on Low level
Sweepmode
SW0
Turn off sweep mode
SW1
Turn on sweep mode
LIN
Linear sweep
LOG
Logarithmic sweep
STT?
Read start frequency
STP?
Read stop frequency
SWT?
Read sweep time
STT.xxx
Programming of start frequency
STP.xxx
Programming of stop frequency
SWT.xxx
Programming of sweep time
DST
Display start frequency
DSP
Display stop frequency
DSW
Display sweep time
Types of modulation
AM0
No amplitude modulation
AM1
Internal AM
AMX
External AM
AMT?
Read modulation depth
AMI.xxx
Programming of modulation depth with internal
AM
AMX.xxx
Programming of modulation depth with external
AM
AMT.xxx
Display modulation depth
FSK1
Activate FSK
FSK0
Deactivate FSK
FK0?
Read carrier frequency
FK1?
Read hop frequency
FK0.xxx
Programming of carrier frequency
FK1.xxx
Programming of hop frequency
PSK0
Deactivate PSK
PSK1
Activate PSK
PH0?
Read phase 0
PH1?
Read phase 1
PH0.xxx
Programming of phase 0
PH1.xxx
Programming of phase 1
Arbitrary mode
ARC4
Initialize 4 K word function, reset value and
address counter
ARC16
Initialize 16 K word function, reset value and
address counter
ARE
Stop loading procedure, calculate intermediate
values
ARP4.:.xxxx=yyyy
4 K word function: write point coordinates to X
and Y.
ARP16:xxxx=yyyy
16 K word function: write point coordinates to X
and Y
ARP4:yyyy=?
4 K word function: read Y coordinate
ARP16:yyyy=?
16 K word function: read Y coordinate
Instrument status
The command "STA?" will cause the instrument to transmit the
following data via the interface:
OT1 RFI SW0 SIN AM0 SK0 CTM VPP
The data fi elds may contain the following data:

Advertisement

Table of Contents
loading

Table of Contents