General information regarding the CE marking HAMEG instruments fulfi ll the regulations of the EMC directive. The conformity test made by HAMEG is based on the actual generic- and product standards. In cases where different limit values are applicable, HAMEG applies the severer standard. For emission the limits for residential, commercial and light industry are applied.
C o n t e n t Declaration of Conformity General information regarding the CE-marking Distortion meter HM8027 Specifi cations Control elements Operational check Ajustment Circuit and layout diagrams Subject to change without notice...
Page 4
H M 8 0 2 7 D i s t o r t i o n M e t e r H M 8 0 2 7 Frequency range 20 Hz to 20 kHz Mainframe HM8001-2 Resolution up to 0.01 % Low inherent distortion of 0.005 % Automatic frequency fine tuning (capture range 15 %) Control output for residual distortion measurement...
2) Connect the output of the HM8037 to the input 100%-CALIBRATION HM8027. (Pushbutton switch) 3) With the above setting the display of the HM8027 should Selection of calibration mode. Adjust ment for 100% reading indicate a value ≥100. with LEVEL Check of input signal attenuation...
Page 7
1) Connect the output of HM8037 to input of HM8027. Check of monitor output Set the frequency to 1 kHz. Setting: 2) Calibrate the HM8027 to 100% and align the frequency by Cal. max. 0dB released means of the control knob...