T R A N S F E R R I N G T H E S A V E D D A T A
Install the NISSEI HealStyle App from the Google Play Store or
¥
Apple App Store to your smartphone, then start the App.
• Compatible devices:
– iPod/iPhone/iPad device: iPhone 4S or iPod touch (5th Generation) or
later with iOS 8 or later
– Android device: Android 4.3 device or later equipped with Bluetooth 4.0
or later.
• Perform the user registration beforehand.
Transfer the saved data.
ø
Touch the Blood Pressure Monitor icon on the App screen.
Press and hold the Bluetooth connection button (
for about 3 seconds.
T R O U B L E S H O O T I N G
Blood pressure is extremely high or low.
• The main unit is not adjusted to your heart. Adjust the main unit to the height of your
heart.
• Measurement is not taken with the correct posture or the cuff is wrapped incorrectly.
Reconfirm the measurement procedures.
• The reading was effected by moving or talking. Remain still and quiet during
measurement.
Measured values vary all the time.
Mental and physical conditions or measurement conditions are influencing blood pressure.
Make measurement under the same conditions.
Measured value is different from that taken at a hospital.
Mental conditions such as nervousness at a hospital or feeling relaxed at home are
influencing blood pressure. Take records of blood pressure taken at home and consult
with your doctor.
Blood pressure is changing 24 hours a day in relation with mental and physical
conditions. Eating, drinking, smoking, exercising or bathing will alter blood pressure.
Blood pressure will be also influenced by mental conditions such as feeling tense or
relieved and health conditions. It is important to know the tendency of blood pressure by
making measurement regularly. Make your own rules including conditions and time to
make measurement and take blood pressure everyday.
Inflation is repeated.
The cuff is inflated again if a sudden change of pressure, which could be due to movement
or straining of the arm or wrist, was detected. Inflation is repeated until blood pressure is
taken. Repeated inflation is not a malfunction of the unit. Do not move or speak during
measurement.
[Err-1] or [Err-3] is displayed.
• Blood pressure could not be taken due to moving or speaking. Do not move or speak
during measurement.
• The result of the measurement is outside of the measurable range. If [Err-1] or [Err-3] is
displayed for each time you measure the blood pressure, consult with your doctor.
P R O D U C T S P E C I F I C A T I O N S
Model
Operating principle
Indicator
Pressure indicating range
Measuring range
Accuracy*
Inflation
Deflation
Exhaust
Power supply
Power consumption
Electric rating
Memory
Bluetooth standard
Operating conditions
Transportation/storage conditions
Cuff
) on the main unit
: DS-S10
: Oscillometric method
: 15-digit liquid crystal display
: 3 to 300 mmHg (cuff pressure)
: 50 to 250 mmHg (systolic), 40 to 180 mmHg (diastolic),
40 to 160 bpm (pulse rate)
: ±3 mmHg (cuff pressure), ±5% of reading (pulse rate)
: Automatic inflation
: Automatic (electric control valve)
: Automatic exhaust valve
: Four 1.5 volt LR6 (AA alkaline) batteries or AC adaptor
ADP-W5 series
: 4 W (max.)
: DC 6V/4 W, with four LR6 batteries
Adaptor; AC 100–240 V, 50/60 Hz, 0.12 A,
Unit; DC 6 V, 500 mA, with designated AC adaptor
: M1/M2; 2 banks, each stores 60 readings
ID (No.1 to No.5); 5 banks, each stores 20 readings
: Bluetooth Low Energy 4.1
: +10˚C to +40˚C, 30% to 85% RH (non condensing)
: –20˚C to +60˚C, 10% to 95% RH (non condensing)
: Coverage arm circumference; 22 to 42 cm, applied part;
BF
The Bluetooth mark starts flashing and the unit is
switched to the connection mode.
NISSEI
HealStyle
Touch the reception button on the upper right on
the App screen, and register the ID for connection
displayed on the unit.
The data communication starts.
• For details about operations of the App, visit: <http://
www.nissei-kk.co.jp>.
• Once the unit is connected with the NISSEI HealStyle App, the clock of the unit is automatically
血圧計
synchronized with the connected smartphone.
• If the Bluetooth connection cannot be established, turn off the unit by pressing the Start/Stop
button (
Do not remove the batteries from the unit or disconnect the AC adapter while transferring the
Press and hold
data.
[Err-2] is displayed.
• The air hose is bent or crushed. Check the air hose.
• The cuff is not securely connected to the main unit. Secure the air connector.
• The cuff is not wrapped appropriately. Reconfirm how to put the cuff on your arm.
• The reading was effected by moving or talking. Remain still and quiet during
measurement.
[Err-300] is displayed.
Over pressure: blood pressure could not be taken due to moving or speaking although the
cuff was inflated to the maximum pressure. Do not move or speak during measurement.
Nothing is displayed.
• The AC adaptor is not connected. Connect the AC adaptor.
• The batteries are depleted. Replace all batteries with new ones.
• The batteries are inserted incorrectly. Reinsert the batteries correctly.
• The battery terminals are not clean. Clean the terminals with a dry cloth.
Clock is not displayed.
The clock is not activated. Adjust the date and time and activate the clock.
• NOTE: The clock is deactivated when the batteries and/or AC adaptor are/is removed.
Measurement date and time are displayed as "- - / - -" and "- - : - -".
The clock is not activated. Adjust the date and time and activate the clock.
Measurement was taken before the clock was activated. The measurement date and time
cannot be saved without the clock being activated.
[0 0] is displayed.
The Start/Stop button (
or when AC adaptor was being plugged in. Turn off the unit once with the Start/Stop
button (
If you cannot get a correct measurement with the methods above, contact your dealer.
Do not disassemble or tamper with the internal mechanism.
Main unit
Protection against harmful ingress of water : IP20
Protection against electric shock
Mode of operation
Classification
Key to symbols
Specifications are subject to change without prior notice.
This product complies with JIST1115 (noninvasive electric blood pressure monitors) and EMC
(electromagnetic compatibility) Standard IEC60601-1-2:2007. (CISPR Classification: Group 1, Class B).
* Test procedure is based on ANSI/AAMI SP10:2002, American National Standard for electronic or
automated Sphygmomanometers, and Amendment.
• iPod, iPhone, iPad are trademarks of Apple Inc., registered in the U.S. and other countries.
• Android is a trademark of Google Inc.
• The Bluetooth
word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and
®
any use of such marks by NIHON SEIMITSU SOKKI CO., LTD. Corporation is under license. Other
trademarks and trade names are those of their respective owners.
− 4−
), then try again.
) was being pressed when the batteries were being installed
).
: Weight: Approx. 331 g (without batteries)
Size: 162.3 x 127.3 x 96 mm (W x D x H)
: Internally powered equipment, Type BF applied part (cuff)
: Continuous operation
: Internally powered equipment
: Type BF applied part
: Refer to instruction manual/booklet.
: Class II equipment
ID FOR
CONNECTION
BLUETOOTH
MARK
受信
測定結果(ニックネーム1)
受信
日付
時刻 最高 最低 脈拍
リスト
グラフ
プロフィール
ID