Make sure that the specified paper (Part Number 36-PSA1600) is
used as excessive wear on the thermal print head may occur if paper
from other manufacturers is used. The following precautions should
be observed when storing or handling the thermal paper rolls:
Do allow the printer to run without paper loaded.
Do not store paper rolls in places exposed to direct sunlight.
Do not use pastes containing organic solvents such as
alcohol, esters or ketones.
Do not allow the paper to be exposed to vinyl chloride.
Do not use cellophane tape on the paper.
Do not allow printer paper to be exposed to diazo copy
paper that has been processed.

Battery Maintenance

The PrinterNOx is powered from a 7.2volt 600mA-hr battery pack
located under the paper cover or from the AC adapter supplied.
The battery pack is fully discharged when shipped from the factory.
Note: Ensure that the dc plug is inserted before the mains supply is
connected to the unit and that the main is switched off before
removing the plug.
Please re-charge the pack with the AC adapter supplied before use.
The battery pack will exhibit a slow decrease in capacity at a rate
dependant upon usage. If the PrinterNOx is normally powered from
the AC adapter then the battery pack will be on continuous trickle
charge. Under these conditions, the capacity will be greater than
60% of nominal capacity for a period of at least 5 years. If the
PrinterNOx is normally powered from the battery pack then the
capacity will be greater than 60% of nominal capacity for at least
1000 charge/discharge cycles. To obtain maximum life, the batteries
29