Connect a NOxSORB to the exhaust of the PrinterNOx, as shown
overleaf, orientated such that the arrow is pointing away from the
instrument. The NOxSORB should be replaced after 24 hours or
when the mauve spheres turn dark brown. When re-ordering please
quote 36-SORB100 for a box of 10 of 36-SORB200 for a box of 20.
Environmental pollution with NO and NO
are present in the connections to the source of NO.

Operation

Before operating the PrinterNOx, ensure that the unit has been fully
charged using the AC adapter supplied. (See Charging Procedure.)
The PrinterNOx may also be operated with the internal rechargeable
batteries.
Connect the water trap with integral hydrophobic PTFE filter and
sampling line to the PrinterNOx as shown below:
Water trap
retaining
screws
Water
Trap
Sampling
Line
Switch on the PrinterNOx. Please note that the water trap must
be used at all times during measurement to avoid
contamination of the fuel cells.
Micro
Alarm
Medical
7
may also arise when leaks
2
PrinterNOx
Exhaust
port
1
2
3
4
5
6
7
8
9
0
PAPER
C AN C EL
C H AR G IN G
EN T ER
NOxSORB