Portable K-Type Fast Response Thermometer for

SPECIFICATIONS

• RANGE*
• METER ACCURACY
@ 23.0 °C ±5 °C
• PROBE ACCURACY ±1.6 ºC (-50.0 to 300 ºC)
FC766C1
*The measurement range may be limited by probe type, and applies to the probe shaft.

OPERATIONAL GUIDE

• To switch ON and OFF: press ON/OFF (
• To take measurements: connect a suitable temperature probe to the meter.
At start-up the thermometer displays all LCD segments for a few seconds,
followed by the percentage indication of the remaining battery life. It then
enters CAL Check mode followed by measurement mode.
If a temperature probe is plugged in, the meter displays the measured
temperature.
If reading is slightly beyond the measurement range, the meter flashes the
closest full-scale value.
To switch the meter OFF, press the ON/OFF key.
• For auto-off feature:
To save battery life, each model is provided with an auto-off feature, which
switches the instrument off after a certain period of non use.
The auto-off feature can be set in 3 modes from SETUP MENU: 8 minutes,
60 minutes or disabled.
• Calibration Check mode:
The meter is provided with an automatic CAL Check feature at start-up, which
allows the user to check the instrument calibration status. If the meter is
calibrated, the display will stabilize and show "0.0 °C" (±0.3 °C) for a
few seconds.
Warning messages:
• "NO PROBE" message: when the probe is not connected or the signal is
interrupted.
• "Err" message: if during Calibration Check the meter displays more than
±0.3 °C. Contact your local Hanna Instruments Office for recalibration.
• DAMAGED PROBE message: when the readings are very far from the
measurement range, the probe may be damaged. The display shows
blinking dashes and "NO PROBE" message is displayed.
Quick Reference Guide
HI9350011
Foodcare Products
-50.0 to 199.9 ºC ; 200 to 300 ºC
-58.0 to 399.9 ºF ; 400 to 572 ºF
±0.4 ºC (-50.0 to 300 ºC)
±0.7 ºF (-58.0 to 572 ºF)
±2.9 ºF (-58.0 to 572 ºF)
).

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI9350011 and is the answer not in the manual?

Questions and answers

Summary of Contents for Hanna Instruments HI9350011

  • Page 1 • “Err“ message: if during Calibration Check the meter displays more than ±0.3 °C. Contact your local Hanna Instruments Office for recalibration. • DAMAGED PROBE message: when the readings are very far from the measurement range, the probe may be damaged. The display shows...
  • Page 2 METER SETUP • To enter Setup MODE: press SET button located inside the battery compartment. SET button • To switch between “°C“ or “°F“: press ON/OFF ( Long press ON/OFF button once to save and continue to auto-off settings. • To cycle through the Auto-Off choices: press ON/OFF ( Long press ON/OFF button once to save and continue to normal operation.

Table of Contents