Curro Mode (Current Output Mode); Curro 20Ma Cal. (Calibration Of Current Output) - BRANNSTROM BilgMon488 Instruction Manual

15 ppm bilge alarm
Table of Contents

Advertisement

Note: Before inspecting the log, make sure the clock is set to the wished time frame (local time or
UTC).
Press
to display the cursor (under leftmost digit).
Press
/
to increase/decrease digit.
Press
to move cursor right one digit (cursor moves
cyclically from left to right, one step per
Save setting by holding
Abort input at any time by pressing
disappears, previous values are filled in).

10.11.5 Curro mode (Current output mode)

The current output of BilgMon488 can be altered between two modes, 0-20 mA and 4-20 mA. These
intervals correspond to 0-30 ppm oil content as displayed in the main menu.
The current output terminal can be used for external printing/logging of the momentary oil content
measurement.
Press
to display the cursor.
Press
/
to alter mode.
Save setting by holding
Abort input at any time by pressing
disappears, previous values are filled in).

10.11.6 Curro 20mA cal. (Calibration of current output)

If current output corresponds badly to the displayed oil content the current output can be calibrated
in this manu (one point calibration).
When entering this menu the BilgMon488 will try to output 20 mA (corresponding to 30 ppm).
Measure the output with ampere meter. If reading is not correct adjust the output according to
below instruction.
Press
to display the cursor.
Press
/
to increase/decrease the current output
signal (adjust to 20 mA output).
Save setting by holding
Abort input at any time by pressing
disappears, previous values are filled in).
).
until cursor disappears.
(cursor
until cursor disappears.
(cursor
until cursor disappears.
(cursor
BRANNSTROM SWEDEN AB
www.brannstrom.se
BilgMon488 Instruction Manual vAD 25(38)
Set clock:
140327 11:34:05
Curro mode
0 to 20 mA
Curro 20mA cal.
59100