Contrast Adjustment; Simulation - BRANNSTROM BilgMon488 Instruction Manual

15 ppm bilge alarm
Table of Contents

Advertisement

10.4 C
ONTRAST ADJUSTMENT
While in main menu press and hold
With
held press
contrast setting by 10%.
10.5 S
IMULATION
The simulation menu lets user simulate the PPM input from the sensor unit as well as the separator
status input state to verify that the installation of the unit is correct and that it responds as expected
to different oil content measurements.
Simulated events will be stored in the log starting with a SIM_ON event and ended with a SIM_OFF
event.
Enter simulation menu
From main menu press and hold
show something similar to the picture on the right.
Keep holding
to stay in the simulation menu.
Leave simulation menu:
Release
.
Toggle the simulated separator status input:
While still holding
separator on/off (
SEP_ON
Increment/decrement simulated PPM value:
While holding
, press
increment/decrement the simulated PPM value in steps
of 1.0.
2
To enter the simulation menu the BilgMon488 needs to be in POWER mode. POWER mode is when
the unit is powered but separator is not running (separator status input is open, see 8.3 Electrical).
.
/
to increment/decrement
2
:
. The LCD will now
, press
to toggle between
/
SEP_OFF
shown in LCD).
/
to
BRANNSTROM SWEDEN AB
www.brannstrom.se
BilgMon488 Instruction Manual vAD 17(38)
Contrast: 010%

Simulation

PPM>30.0 SEP_OFF