Chauvin Arnoux MTX 3290 User Manual
Chauvin Arnoux MTX 3290 User Manual

Chauvin Arnoux MTX 3290 User Manual

Calibration set, asyc iv

Advertisement

Quick Links

C.A 527x
CALIBRATION SET
MX 50xx
KOMETEC
Karl Oelkers e.K.
Mess- und Prüfgeräte · Shop
Mozartstr. 10 · D-88097 Eriskirch
T: 07541 / 955-1313 · F: 07541 / 955-1131
info@kometec.de · www.kometec.de
MTX 3290
MTX 3291
ASYC IV
ENGLISH
User's manual

Advertisement

Table of Contents
loading

Summary of Contents for Chauvin Arnoux MTX 3290

  • Page 1 C.A 527x CALIBRATION SET MX 50xx KOMETEC Karl Oelkers e.K. Mess- und Prüfgeräte · Shop Mozartstr. 10 · D-88097 Eriskirch T: 07541 / 955-1313 · F: 07541 / 955-1131 info@kometec.de · www.kometec.de MTX 3290 MTX 3291 ASYC IV ENGLISH User’s manual...
  • Page 2: Table Of Contents

    English CONTENTS CALIBRATION KIT ...................... 3 1.1 G CA 527 ................. 3 ETTING STARTED WITH 1.1.1 Step 1 ......................... 3 1.1.2 Step 2 ......................... 3 1.1.3 Step 3 ......................... 3 1.2. G MX 5006 ................4 ETTING STARTED WITH 1.2.1 Step 1 ........................4 1.2.2 Step 2 ........................
  • Page 3: Calibration Kit

    English 1. CALIBRATION KIT Equipment required: Calibration kit ref: P01196770 including software and interface board Screwdrivers: Torx 4 mm or Posidriv, depending on date of manufacture Calibrator: Fluke 5500, Datron 9900 or equivalent Test leads Short cord or short circuit with 19 mm spacing 1.1 Getting started with CA 527x Unplug the test leads.
  • Page 4: Getting Started With Mx 5006

    English 1.2. Getting started with MX 5006 Disconnect the power supply cord and test leads. 1.2.1 Step 1 Remove the rubber feet in order to access the screws. 1.2.2 Step 2 Unscrew and separate both sides. 1.2.3 Step 3 Keep the interface board away from the power supply board (risk of electrical shock). Position the interface board as shown below.
  • Page 5: Getting Started With Mx 5060

    English 1.3 Getting started with MX 5060 1.4 Getting started with MTX 329x * Use an USB optic cable * Connect the cable to the MTX 329x.
  • Page 6: Running The Calibration Program

    English RUNNING THE CALIBRATION PROGRAM 2.1. Installing the program Insert the CD into the PC. Run “SETUP.EXE”. 2.2. First run Run MTX_REG_MON program. Follow the instructions. Press “OK”. Select the language: English...
  • Page 7 English Select the instrument range: If CAxxx range, select the model:...
  • Page 8 English Select CA 5271 / 73, if needed: If MXxxxx range, select the model: If MTXxxxx range, select the model: E²PROM read of the CA 527x / MX 50xx / MXT 329x : Select "NO".
  • Page 9: Com Port

    English 2.3. COM port Select the communication port: look for the COM port that corresponds to your USB cable. Example: Windows XP config panel System Hardware Device manager Ports (COM and CPT). Search for CP210x in the list of COM ports. Select the corresponding COM port in menu below.
  • Page 10 English...
  • Page 11: Calibration

    English 3. CALIBRATION The CA 527x, MX 50xx, MTX 329X have manual settings. 3.1. Offset setting Warning! The operations below must be carried out strictly in the following order. 1) Turn the multimeter OFF. 2) Short-circuit the VOLT and COM inputs. 3) Press the "MODE"...
  • Page 12 English 1) Select (if this has not already been done previously) a function and its range Ex.: VDC 600mV. 2) Press the yellow button "CAL Pentes / Slopes". This opens the following menu. 3) Calibrate. Note: R0 to R5 are the different ranges. The boxes to the right show the relevant range for each instrument.
  • Page 13 English Loop : Check whether the "Input" set point is correct (e.g. range 600 mV =+500mV). Apply the set point positive voltage. Press the "READ val AD" (Input pos) button. Apply the set point negative voltage. Press the "READ val AD" (Input neg) button. If the calculated slopes are correct (type table below), press the "WR Slope / Slope".
  • Page 14 English Typical Offsets and Slopes for CA 527X Typical Offsets and Slopes for MX 50xx Func Offset+-10% Pente+-5% Func Offset+-10% Pente env. Volt 60mV + 16384 Volt 60mV + 61440 Volt 60mV - 16384 Volt 60mV - 61440 Volt 600mV + 16384 0.375 Volt 600mV +...
  • Page 15 English Func Offset+-10% Pente env. Volt 60mV + 61440 Volt 60mV - 61440 Volt 600mV + 61440 0.375 Volt 600mV - 61440 0.375 Volt 6V + 61440 Volt 6V - 61440 Volt 60V + 61440 3.45 Volt 60V - 61440 3.45 Volt 600V + 61440...
  • Page 16: Correction Factor

    English 3.3. Correction factor For use only by SAV 1) In the "Function MENU" select the "Cor_AC" line. 2) Check whether the correction slopes have been refreshed. 3) Carry out a transfer using the "WR Slope/Pente" button. 3.4. Bandwidth Compensation To calibrate the bandwidth, select: 1) Menu "Fr_Comp", "AC + DC"...
  • Page 17 English 4) At the end of the process, save the setting with the command "MEM_RAM" or "Write to EEPROM mem." 5) Note: The program provides a starting configuration with all Cx OFF. After the initialization, switching ON any Cx can only decrease the value. It is possible to start with positions corresponding to the preceeding adjustment.
  • Page 18: Misc. (Divers)

    English 3.5. Misc. (Divers) Select the language Store / read the calibration data to / from a file and transfer to the CA 527x / MX50xx / MTX 329x or to a computer. 3.6. READ Here you can read some Multimeter data : Read_LCD reads the value of the LCD primary display Read_UC reads the RMS value supplied by the converter Read_Bat reads the battery voltage...
  • Page 19: Display Offsets And Slopes

    English 3.8. Display Offsets and Slopes Press this key. Ex : MX 5060 3.9. Running a remote command To pilot the CA 527x, the MX 50xx or the MTX 329x manually, select a command in the list below and press the "Command "...
  • Page 20: Storing Calibration Data

    English 4. STORING CALIBRATION DATA Select the "Write mem to EEprom" command and press the "Command " button. The calibration results are stored in the EEprom. 5. QUITTING THE PROGRAM You can save your modifications on the EEprom : When exiting the program, the language and the COM settings are memorized.