Emergency Firmware Update; Factory Default - CEL-MAR ADA-7010D User Manual

Multimode fibre-optic to rs-232 digital converter
Table of Contents

Advertisement

ADA-7010D
During loading software the yellow LED located beside SW1 micro-switch will blink, showing data flow to the converter. If the software
was loaded correctly yellow LED will be blink with frequency 1 Hz.
After that set microswitch SW1 to run mode as shown in the table below.
SW1-1
SW1-2
OFF
OFF
Yellow LED will be OFF.

6.6. EMERGENCY FIRMWARE UPDATE

In case of the unsuccessful update of the converter software, try again according to point FIRMWARE UPDATE. If the update is still
incorrect use emergency firmware update. Set SW1 microswitch mode as in the table below.
SW1-1
SW1-2
ON
ON
After micro-switch setting, ADA-7010D should be restarted, by turning OFF and then ON the power supply. The yellow LED will light
continuously and the converter will be in Emergency Firmware Update mode. Now follow the description in point FIRMWARE
UPDATE.
After successful updating, set SW1 micro-switch as in table below. Yellow LED will be OFF.
SW1-1
SW1-2
OFF
OFF

6.7. FACTORY DEFAULT

In case of faulty functioning of ADA-7010D, like:
- no communication in the configuration mode,
- transmission errors,
can be restored the factory default setting of the converter internal registers.
Factory default
Parameter
Baud rate
Data bits
Parity
Stop bits
Fig. 8. Selection of firmware file
RS232
Fibre-Optic
interface
interface
9600bps
460800 bps
8
Non
1
10
8
Non
1