Power connection
• Power requirements: The display must be powered using the
supplied medical approved 24 VDC power supply.
• The medical approved DC power supply must be powered by the
AC mains voltage.
• The display complies to the mentioned standards only when
used with the supplied medical power supply.
• The display must be powered from a center-tapped electrical
circuit when used in the USA at voltages above 120 Volts.
• The display is intended for continuous operation.
• The display is powered from an external power supply for class I
equipment. The installer is responsible for testing the display's
earth ground to verify that it meets the impedance
requirements for the given country regulatory requirements.
• The display must be earthed by a minimum 18 AWG, maximum
6 foot (1.8m) long wire connected to the ground connection at
the rear.
Power cords:
• Utilize a UL-listed detachable power cord, 3-wire, type SJ or
equivalent, 18 AWG min., rated 300 V min., provided with a
hospital-grade type plug 5-15P configuration for 120V
application, or 6-15P for 240V application.
• Do not overload wall outlets and extension cords as this may
result in fire or electric shock.
• Mains lead protection (U.S.: Power cord): Power cords should be
routed so that they are not likely to be walked upon or pinched
by items placed upon or against them, paying particular
attention to cords at plugs and receptacles.