Ecom Instruments MCAL 4200 Instruction Manual page 87

Table of Contents

Advertisement

Available languages

Available languages

Read (mV)
Source (mV)
Maximum current output in voltage ranges is 1mA with an output impedance of
TC Type
J
K
T
E
R
S
B
C
XK
BP
L
U
N
All TC errors include CJC errors
CJC errors outside of 23 ± 5 °C is 0.05°C/°C (In °C add .2 for cold junction compensation error.)
86
Table 15: Thermocouple Measurement / Source
Range
-10.000mV - 75.000mV
-10.000mV - 75.000mV
Table 16: Thermocouple Read and Source
-210.0 ... 0.0
800.0 ... 1200.0
-200.0 ... 0.0
0.0 ... 1000.0
1000.0 ... 1372.0
-250.0 ... 0.0
-250.0 ... -100.0
-100.0 ... 1000.0
0.0 ... 1767.0
0.0 ... 1767.0
600.0 ... 800.0
800.0 ... 1000.0
1000.0 ... 1820.0
0.0 ... 1000.0
1000.0 ... 2316.0
-200.0 ... 800.0
800.0 ... 2500.0
-200.0 ... 0.0
-200.0 ... 0.0
-200.0 ... 0.0
0.0 ... 1300.0
Accuracy(% of reading ± digit)
0.02% ± 10
0.02% ± 10
Range (°C)
0.0 ... 800.0
0.0 ... 400.0
0.0 ... 800.0
0.0 ... 900.0
0.0 ... 600.0
< -
Accuracy (°C)
0.4
0.2
0.3
0.6
0.3
0.5
0.6
0.2
0.6
0.2
1.2
1.2
1.2
1.3
1.5
0.6
2.3
0.2
0.9
2.3
0.25
0.2
0.5
0.25
0.8
0.4