Specifications
The following specifications are typical for the full operating temperature range unless otherwise
noted. Refer to the
Environmental
Simultaneous Analog Input
Number of channels.......................................... 16
ADC resolution................................................. 12 bits
Input range
Typical ...................................................... ±5 V, ±10 V
Minimum .................................................. ±4.95 V, ±9.90 V
Common-mode range ....................................... ±10 V
Sample rate (per channel) ................................. 100 kS/s max
Accuracy
Nominal
Range (V)
Measurement Conditions
±5
Maximum (-40 °C to 85 °C)
Typical (23 ° ±5 °C)
±10
Maximum (-40 °C to 85 °C)
Typical (23 ° ±5 °C)
AbsoluteAccuracy = Reading(GainError) + Range(OffsetError) + Noise
* Local ambient temperature. Refer to the
temperatures.
†
Offset error includes the effect of INL
Stability
Gain drift................................................... 15 ppm/°C
Offset drift ................................................ 15 µV/°C
Noise................................................................. 1.5 mV
-3 dB bandwidth ............................................... 210 kHz
CMRR (f
= 60 Hz).......................................... 60 dB min
in
Input impedance
Differential................................................ 240 kΩ
Single-ended ............................................. 120 kΩ
Overvoltage protection ..................................... ±30 V max
section for more information on operating temperatures.
*
Environmental
.
NI 9683 User Guide and Specifications | © National Instruments | 17
Percent of
Reading
(Gain Error)
0.70%
0.25%
0.70%
0.25%
section for more information about operating
rms
Percent of
Range
†
(Offset Error)
0.28%
0.12%
0.16%
0.07%