Agilent Technologies E6000C User Manual page 258

Mini-otdr
Hide thumbs Also See for E6000C:
Table of Contents

Advertisement

Performance Test Form Sheets
Performance Test for the Agilent E6000C with Single-mode Modules
Model Agt. __________ Module Report No. __________
No. Test Description
III. Attenuation Deadzone 1300 nm Wavelength
Return Loss ≥ 35dB
Conditions:
Meas. Span. 2km
Pulsewidth 10ns
IV.
Distance Accuracy 1300 nm Wavelength (Optional test)
Fiber Length: ________ m
Start Position: ________ m
|Distance Accuracy| = (Fiber Length x Scale Error + Offset Error + 1/2 Sample Spacing)
|Distance Accuracy| = (_______ m x 10
|Distance Accuracy| = ___________ m
Minimum Distance = Fiber Length - | Distance Accuracy|
Maximum Distance = Fiber Length +| Distance Accuracy|
Meas.Span
Pulsewidth
2 km
1 µ s
100 ns
E6000C Mini-OTDR User's Guide, E0302
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
D Single-Mode/Multimode Module Performance Tests
Maximum Specification
Attenuation
Deadzone
m
_______
Sample Spacing: ________ m (as ∆ on the screen)
-4
+1m
Minimum
Measured
Distance (typ.)
Distance
m
m
_______
_______
_______
Page 6 of 6
Date ___________
Meas.
Uncertainty
m
m
10
_______
+ ________ m)
Maximum
Meas.
Distance (typ.)
Uncertainty
m
m
_______
_______
257

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents