D Single-Mode/Multimode Module Performance Tests
Performance Test for the Agilent E6000C with Single-mode Modules
Model Agt. __________ Module
No. Test Description
I.
Dynamic Range 1310 nm Wavelength
Pulsewidth
Dynamic
Range
98%
dB
10 µ s
_______
1 µ s
_______
100 ns
_______
10 ns
_______
II.
Event Deadzone 1310 nm Wavelength
Return Loss ≥ 35dB
Conditions:
Meas. Span. 2km
Pulsewidth 10ns
E6001A, E6003A,
E6003B, E6004A,
E6008B, E6013A
246
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Report No. __________ Date ___________
Minimum Specification
Dyn Range =
E6001A
E6003A E6003B E6008B E6013A
Dyn Range
E6004A
98%
+ 1.9dB
dB
dB
dB
_______
28
35
_______
23
30
_______
18
24
_______
13
19
Event Deadzone
_______ m
E6000C Mini-OTDR User's Guide, E0302
Performance Test Form Sheets
Page 3 of 8
Uncertainty
dB
dB
dB
38
42
36
30
35
29
24
29
23
19
24
18
Max Spec
Uncertainty
5 m
_______ m
Meas.
dB
_______
_______
_______
_______
Meas.

Hide quick links: