D Single-Mode/Multimode Module Performance Tests
Settings
N O TE
Configuration
224
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
3 Set the OTDR:
[S
] menu:
ETTINGS
– <R
> - select <R
ANGE
Start: - enter value ST from Table 2, Table 3, or
Table 4. Confirm with <O
Span: - enter value SP from Table 2, Table 3, or Table
4. Confirm with <O
– <P
W
>: enter value PW from Table 2, Table
ULSE
IDTH
3, or Table 4.
– <W
>: If a dual wavelength module is
AVELENGTH
installed, select the required wavelength
– <M
. M
>: Averaging
EAS
ODE
– <O
M
>: Dynamic
PTIMIZE
ODE
– <A
. T
>: 3 min
VG
IME
If the averaging parameter is listed for Number of
Averages, you should do the following:
• Exit the [S
] menu
ETTINGS
Press Ok.
• Enter the Instrument Config screen.
Select [C
.]<I
ONFIG
NSTRUMENT
• Bring up the OTDR Settings page
Select [P
I
]<OTDR S
AGE
NDEX
• Select Averaging time
Move to the Averaging Mode box and press select,
select Averaging time from the menu you see.
• Save this configuration
Select Save.
• Exit the Instrument Config screen
Select Ok.
E6000C Mini-OTDR User's Guide, E0302
Test I. Dynamic Range
I
...>:
ANGE
NPUT
>.
K
>.
K
C
>
ONFIG
>
ETTINGS