D Single-Mode/Multimode Module Performance Tests
Performance Test for the Agilent E6000C with Single-mode Modules
Model Agt. __________ Module
No. Test Description
I.
Dynamic Range 1550 nm Wavelength
Pulsewidth
Dynamic
Range
98%
dB
10 µ s
_______
1 µ s
_______
100 ns
_______
10 ns
_______
II.
Event Deadzone 1550 nm Wavelength
Return Loss ≥ 35dB
Conditions:
Meas. Span. 2km
Pulsewidth 10ns
E6003A, E6003B,
E6004A, E6008B,
E6012A, E6013A
248
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Report No. __________ Date ___________
Minimum Specification
Dyn Range =
E6003A E6003B E6004A E6008B
Dyn Range
98%
+ 1.9dB
dB
dB
dB
_______
34
37
_______
29
29
_______
22
22
_______
17
17
Event Deadzone
_______ m
E6000C Mini-OTDR User's Guide, E0302
Performance Test Form Sheets
Page 5 of 8
E6013A
E6012A
Uncertainty
dB
dB
dB
28
41
35
23
34
28
18
27
22
13
22
17
Max Spec
Uncertainty
5 m
_______ m
Meas.
dB
_______
_______
_______
_______
Meas.